GUI reference
R&S
®
ZNA
815
User Manual 1178.6462.02 ─ 20
Note:
Remote languages other than DEFAULT are intended for remote control of the
analyzer. A mixed approach, with part of the instrument configuration defined via the
GUI, is possible but can cause unexpected results in some instances.
Remote command:
Define *IDN + *OPT...
Defines a format for the ID string and the option string of the analyzer. The default
strings are automatically adjusted to the selected "Remote Language". The strings can
be queried via
*IDN?
and
*OPT?
, respectively.
●
If the DEFAULT language is activated, the factory ID string
Rohde&Schwarz,ZNA<Max. Freq>-<Ports>Port,<Order and Serial No>,<FW_Ver-
sion>
(e.g.
Rohde-Schwarz,ZNA26-4Port,1332450024100005,1.50.2.19
) is set.
The option string is a comma-separated list of all installed software and hardware
options. The bit order for transferred binary data is swapped (
FORMat:BORDer
SWAPped
).
●
If the PNA language is activated, Keysight-compatible ID and option strings are
set. The bit order for transferred binary data is normal.
●
If one of the "HP xxxx" languages is activated, "HP xxxx"-compatible ID and option
strings are set. Binary data is transferred in a device-specific bit order, however,
the bit order can be changed using "HP xxxx"-specific commands.
The ID and option strings can be changed or reset to the R&S factory ID string.
Remote command:
SYSTem:IDENtify[:STRing]
SYSTem:IDENtify:FACTory
SYSTem:OPTions[:STRing]
SYSTem:OPTions:FACTory
FORMat:BORDer
Advanced ...
Opens a dialog that allows to set the following parameters:
●
"Wait for Data after Sweep" determines the execution behavior of
INITiate[:IMMediate]
commands (see
Chapter 7.3.8, "INITiate commands"
on page 1170).
If enabled, an automatic
*WAI
Chapter 7.2, "Common commands"
on page 882).
By default this is disabled.
Setup softtool