Test Settings
Allows you to set the absolute or relative number of mask violations that will result in a
mask fail condition.
Remote command:
CALCulate<Chn>:EYE:MASK:FAIL:CONDition
CALCulate<Chn>:EYE:MASK:VIOLation:TOLerance
CALCulate<Chn>:EYE:MASK:VIOLation:RATE
Save / Load Mask Configuration
Opens a dialog that allows to save/load the mask test configuration to/from a 7bit
ASCII file (
*.mask
). An example is shown below:
[Mask Top]
Active = true
Width Top = 9e-009
Offset Top = 6
[Mask Bottom]
Active = true
Width Bottom = 9e-009
Offset Bottom = 6
[Mask Polygon]
Active = true
Shape = Octagon
Main Width = 9e-009
Minor Width = 8e-009
Main Height = 9
Minor Height = 8
[MASK CENTER]
Horizontal = 0
Vertical = 0
[TEST SETTING]
Fail Condition = Samples (Percent)
Violation Tolerance = 1
Violation Percent = 10
Remote command:
MMEMory:LOAD:EYE:MASK
MMEMory:STORe:EYE:MASK
Automatic Mask Generation
The "Automatic Mask Generation" function attempts to size the polygon, top, and bot-
tom masks in a reasonable manner based on the current eye measurement settings.
Remote command: