GUI reference
R&S
®
ZNA
553
User Manual 1178.6462.02 ─ 20
●
For an automatic calibration (left), it displays the port assignment.
●
For a manual calibration (right), it displays the ports with their connector, gender
and cal kit.
For a given port, these values are shared among the related manual channel cali-
bration definitions.
The buttons "Same Connector all Ports", "Same Gender all Ports" and "Import Cal
Kit..." provide the functionality known from the
Start Cal
Saves the calibration definitions to the current setup. Then enters the
to acquire data for all involved channels, ports and standards.
Note:
In the current implementation, "Start Cal" is disabled if both manual and auto-
matic calibration definitions are enabled in the
.
Save/Cancel
Saves the channel calibration definitions to the current setup or discards unsaved mod-
ifications.
5.11.1.7
Guided Calibration wizard
The "Guided Calibration" wizard lets you calibrate a set of channels using the calibra-
tion methods the firmware proposes for the related measurements.
If a calibration unit is connected, it is automatically used.
Access:
Channel – [Cal] > "Start Cal" > "Start... (Guided)"
Step 1: channel selection
See
Chapter 5.11.1.8, "Channel Selection for Calibration dialog"
channel selection is shown, even if the active recall set contains only one channel.
Step 2: calibration settings
Defines power settings that apply to all subsequent calibrations and allows you to
select some optional calibration steps.
Cal softtool