GUI reference
R&S
®
ZNA
523
User Manual 1178.6462.02 ─ 20
Incompatibilities between the selected calibration type, the standards and the channel
settings can cause the calibration to be inaccurate. The analyzer auto-detects potential
sources of errors and displays appropriate, self-explanatory notice boxes.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:SAVE:SELected[:DUMMy]
[SENSe<Ch>:]CORRection:COLLect:SAVE:SELected:DEFault
[SENSe<Ch>:]CORRection:COLLect:DELete
[SENSe<Chn>:]CORRection:SSTate?
[SENSe<Ch>:]CORRection:STIMulus?
[SENSe<Ch>:]CORRection:STIMulus:PORT<PhyPt>?
Automatic calibrations
The calibration wizard can also guide you through a calibration using a calibration unit
(plus a power meter for SMARTer Cals).
Background information
●
Chapter 4.5.5, "Automatic calibration"
●
1. For each port assignment:
a) Connect the calibration unit to the related set of test ports.
b) Perform an automatic system error correction.
The upper part of the "Cal Unit" screen displays the calibration sweep diagrams for
the currently measured S-parameter. The lower part visualizes the active port
assignment and the measurement progress.
Cal softtool