Concepts and features
R&S
®
ZNA
270
User Manual 1178.6462.02 ─ 20
4.7.15
In-situ de-embedding
Option R&S
ZNA-K220
This option offers a pre-installed or service-retrofitted version of AtaiTec's
In Situ De-
Embedding
(ISD) tool for fixture de-embedding.
The tool is integrated into the de-embedding functionality of the analyzer firmware.
●
For general information about fixture de-embedding, see
●
For detailed information about the ISD tool, see the product pages at
Tips for using ISD
The ISD User's Guide offers the following guidelines to get the most out of ISD:
1. The Touchstone files must have enough number of points.
The minimum required is at least two points for every 90 degree phase angle
change in insertion loss (or 8 points per wavelength).
2. The "2x Thru’s" return loss must be smaller than its insertion loss.
3. The "2x Thru’s" insertion loss must be relatively linear, with a few wavelengths at
the highest frequency.
4. Use a 4-port "2x Thru" if strong far-end crosstalk (FEXT) is present.
5. Use a
complete
DUT w. fixture file for extraction.
Source: "In Situ De-Embedding (ISD) User's Guide" – November 11, 2018
4.7.16
Smart fixture de-embedding
Option R&S
ZNA-K230
This option offers a pre-installed or service-retrofitted version of PacketMicro's
Smart
Fixture De-embedding
(SFD) for fixture de-embedding.
The tool is integrated into the de-embedding functionality of the analyzer firmware.
●
For general information about fixture de-embedding, see
●
For detailed information about the SFD tool, see the product pages at
www.packetmicro.com/Products/sfd-tool.html
.