Command reference
R&S
®
ZNA
1366
User Manual 1178.6462.02 ─ 20
Suffix:
<Ch>
.
Parameters for setting and query:
<Gain>
Gain factor for the internal low noise amplifier in dB
Range:
20 to 30
Increment: 5
Options:
R&S
ZNA-B302
Manual operation:
See
7.3.14.23
[SENSe:]PULSe...
Pulsed measurements.
[SENSe<Ch>:]PULSe:ATIMe
<Acquisition time>
Defines the duration of the data acquisition for pulse measurements in channel <Ch>.
Suffix:
<Ch>
.
Channel number
Parameters:
<Acquisition time>
Data acquisition time
Default unit: s
Manual operation:
See
[SENSe<Ch>:]PULSe:COUPled[:STATe]
<Boolean>
Enforces identical pulse generator settings for all ports.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean>
*RST:
OFF (0)
Options:
R&S
ZNA<Freq>-B4<Pt> or R&S
ZNA-B91
Manual operation:
See
[SENSe<Ch>:]PULSe:DUTYcycle
<Percentage>
Defines the duty cycle of the generated pulses.
This setting only takes effect, if:
●
Internal Pulse Control is enabled (
[SENSe<Ch>:]PULSe:INTernal[:STATe]
●
Fixed Duty Cycle is enabled (
[SENSe<Ch>:]PULSe:FXDCycle[:STATe] ON
Note
that currently the exact value of the duty cycle can not be guaranteed. However,
you can be sure that the specified value is not exceeded.
SCPI command reference