Remote Control
R&S
®
FSPN
397
User Manual 1179.4363.02 ─ 01
<t>
1..n
irrelevant
Parameters:
<Position>
Percentage of the diagram height with 100
% corresponding to
the upper diagram border
*RST:
50
Default unit: PCT
Example:
//Shift reference position to top of the diagram
DISP:TRAC:Y:RPOS 100
Manual operation:
See
"Scaling the y-axis of the phase diagram"
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RVALue
<RefValue>
This command defines the value assigned to the reference position.
Prerequisites for this command
●
Phase result display must be available and selected.
Suffix:
<n>
.
1..n
<t>
1..n
irrelevant
Parameters:
<RefValue>
<numeric value>
*RST:
0
Default unit: Depends on the selected unit (deg or rad)
Example:
//Define reference value in rad
DISP:TRAC:Y:UNIT RAD
DISP:TRAC:Y:RVAL 5
Manual operation:
See
"Scaling the y-axis of the phase diagram"
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:UNIT
<YScalUnit>
This command selects the unit in which the phase results are displayed.
Prerequisites for this command
●
Phase result display must be available and selected.
Suffix:
<n>
.
1..n
<t>
1..n
irrelevant
Common Measurement Settings