Remote Control
R&S
®
FSPN
380
User Manual 1179.4363.02 ─ 01
VSUPply
Selects the V
supply
connector.
VTUNe
Selects the V
tune
connector.
*RST:
VSUPply
Example:
//Measure the V
aux
connector in the Supply Current / Voltage
result.
CONF:VCO:FIX:SOUR VAUX
CONFigure:VCO:SWEep:DELay:INITial
<SettlingTime>
This command defines an initial delay time that allows the DUT to settle before the first
measurement starts.
Parameters:
<SettlingTime>
<numeric value>
*RST:
10 ms
Default unit: s
Example:
//Define an initial settling time of a 100 ms.
CONF:VCO:SWE:DEL:INIT 100MS
Manual operation:
See
CONFigure:VCO:SWEep:DELay[:POINt]
<SettlingTime>
This command defines a delay time that allows the DUT to settle when a new tuning
voltage is applied and before the measurement starts.
Parameters:
<SettlingTime>
<numeric value>
*RST:
10 ms
Default unit: s
Example:
//Define a settling time of a 100 ms.
CONF:VCO:SWE:DEL 100MS
Manual operation:
See
CONFigure:VCO:SWEep:FCOunter:RESolution
<Frequency>
This command selects the frequency resolution for VCO characterization.
Parameters:
<Frequency>
<numeric value>
If you select a frequency resolution that is not supported, the
next available frequency resolution is selected.
*RST:
1 kHz
Default unit: Hz
Common Measurement Settings