Command reference
R&S
®
ZNB/ZNBT
952
User Manual 1173.9163.02 ─ 62
CALCulate<Chn>:MARKer<Mk>:FUNCtion:SPAN
Sets the span of the sweep range equal to the absolute value of the first coordinate of
the active delta marker
<Mk>
on trace no.
<Chn>
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk>
Marker number.
Example:
*RST; :CALC:MARK ON; MARK:DELTa ON
Create marker 1 in the center of the current sweep range and
enable the delta mode.
CALC:MARK:X 300MHz
Increase the stimulus value of the delta marker by 300 MHz.
CALC:MARK:FUNC:SPAN
Set the sweep range equal to 300 MHz. The sweep range starts
at the reference marker position, i.e. in the center of the analyz-
er's frequency range.
Usage:
Event
Manual operation:
See
"Center = Marker / Start = Marker / Stop = Marker / Span =
CALCulate<Chn>:MARKer<Mk>:FUNCtion:STARt
CALCulate<Chn>:MARKer<Mk>:FUNCtion:STOP
CALCulate<Chn>:MARKer<Mk>:REFerence:FUNCtion:STARt
CALCulate<Chn>:MARKer<Mk>:REFerence:FUNCtion:STOP
These commands set the
STARt
or
STOP
of the sweep range to the stimulus value of
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk>
Marker number (ignored for reference markers)
Example:
*RST; :CALC:MARK ON
Create marker 1 in the center of the current sweep range and
assign it to trace no. 1.
CALC:MARK:FUNC:STAR
Divide the sweep range in half, starting at the current marker
position. As an alternative:
CALC:MARK:FUNC:STOP
Divide the sweep range in half, ending at the current marker
position.
Usage:
Event
Manual operation:
See
"Center = Marker / Start = Marker / Stop = Marker / Span =
SCPI command reference