GUI reference
R&S
®
ZNB/ZNBT
479
User Manual 1173.9163.02 ─ 62
SOURce<Ch>:POWer:CORRection:DATA:PORT<PhyPt>
[SENSe<Chn>:]CORRection:PSTate?
Power Cal dialog – Ref. Receiver
In Ref. Receiver mode, the "Power Cal" dialog guides you through a reference receiver
calibration.
The selected source port is displayed with its cal power settings (see
on page 527). A circuit diagram visualizes the connectivity.
A power meter must be connected to the calibrated source port.
Access:
Background information
Refer to
Chapter 4.5.6.1, "Source power calibration"
Start Cal Sweep
Starts the calibration sweeps for the selected port and power calibration settings and
closes the dialog. The calibration is performed as described in
Open the
if you wish to modify the calibration procedure.
Remote command:
SOURce<Ch>:POWer:CORRection[:ACQuire]
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect[:ACQuire]
Cal softtool