Concepts and features
R&S
®
ZNB/ZNBT
179
User Manual 1173.9163.02 ─ 62
Figure 4-10: Sliding Match: GUI representation
A calibration is valid (and can be applied to the calibrated channel) if either the Match
or three positions of the Sliding Match have been measured. However, it is often desir-
able to acquire calibration data from both standards.
The R&S
ZNB/ZNBT can acquire correction data for up to 20 positions.
The analyzer combines the data in an appropriate manner:
●
The Match results are used up to the lower edge of the specified frequency range
of the Sliding Match (Min Freq).
●
The Sliding Match results are used for frequencies above the Min Freq. In general,
the Sliding Match provides better results than the Match within its specified fre-
quency range.
4.5.2.4
Cal kit Files
Calibration kit files can be used to store the parameters of a particular calibration kit, to
reload the data and to exchange calibration kits from one network analyzer to another.
Cal kit file contents
Cal kit files are independent of the current recall set and contain the following informa-
tion:
●
Name and label of the calibration kit
●
Connector type including all connector type parameters (name, polarity, offset
model, reference impedance)
●
Type, gender and label of all standards in the kit together with the circuit model
parameters (offsets, load) or S-parameter tables (.s<n>p file) that are necessary to
determine its magnitude and phase response.
Calibration