Command reference
R&S
®
ZNB/ZNBT
1458
User Manual 1173.9163.02 ─ 62
Setting parameters:
<ID>
ID of the utilization you want to delete.
Example:
See
DIAGnostic:HUMS:UTILization:CUSTom:ADD
Usage:
Setting only
DIAGnostic:HUMS:UTILization:CUSTom:DELete:ALL
Deletes all custom utilizations.
Example:
See
DIAGnostic:HUMS:UTILization:CUSTom:ADD
Usage:
Event
DIAGnostic:HUMS:UTILization:CUSTom:UPDate
<ID>, <Value>, <Mode>[, <State>]
Changes the current values of the custom utilization (and indirectly also the start
value).
Setting parameters:
<ID>
Identifier of the custom utilization.
<Value>
Defines the value to be changed.
<Mode>
ABSolute
Defines an absolute value for the utilization.
For example: Currently, the utilization value = 5. If you define an
absolute value of 1, the new value is 1.
INCRement
Increases or decreases the current value by a certain amount.
For example: Currently, the utilization value = 5. If you increment
it by 1, the new value is 6.
If you you specify a negative number, it decreases the current
value.
In both cases, the previous value becomes the new start value.
<State>
ON | OFF | 1 | 0
Example:
See
DIAGnostic:HUMS:UTILization:CUSTom:ADD
Usage:
Setting only
DIAGnostic:HUMS:UTILization:HISTory?
[<StartDate>[, <EndDate>[,
<TimeResolution>]]]
Queries the history of utilization activity on the connected instrument.
The data format is either
XML
or
JSON
, depending on
Timestamps follow the Unix time format.