Command reference
R&S
®
ZNB/ZNBT
1226
User Manual 1173.9163.02 ─ 62
–
A VNA cal kit file name without path but prefixed with "SD:"
refers to a specific cal kit file stored on the SD card inserted at
the active calibration unit.
–
A VNA cal kit file name
*.calkit
with path refers to a spe-
cific cal kit file stored in an arbitrary directory on the analyzer.
<TestPort1>
Test port numbers. For an n-port automatic calibration, n arbi-
trary (not necessarily consecutive) port numbers must be speci-
fied.
<TestPort2>
Second test port number of the analyzer
Example:
CORR:COLL:AUTO:TYPE FNPort, '', 1, 2, 4
Perform an automatic full 3-port calibration at test ports 1, 2,
and 4 using the calibration unit's default calibration kit file and
arbitrary test ports of the cal unit.
Usage:
Setting only
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:UTHRough
<TestPort1>, <TestPort2>
With this command, you can perform the Unknown Through measurement during an
automatic UOSM calibration.
Although it works for all calibration units, it is only useful if the calibration unit selected
using
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess
(ICS).
Suffix:
<Ch>
.
Channel number
Setting parameters:
<TestPort1>,
<TestPort2>
VNA test port numbers
Example:
See
Chapter 8.2.5.5, "Inline calibration"
Usage:
Setting only
[SENSe:]CORRection:COLLect:AVERage
<Average>
Activates automatic averaging, which means that the VNA may perform multiple cali-
bration sweeps and apply averaging to reduce trace noise. In contrast to manual aver-
aging (see
) the number of calibration sweeps is
calculated automatically.
Parameters:
<Average>
AUTO | MANual
Manual operation:
See
SCPI command reference