Command reference
R&S
®
ZNB/ZNBT
1086
User Manual 1173.9163.02 ─ 62
Note:
*RST
or "Preset" do not change the configuration of the
to restore
default values.
*RST:
n/a (default: NOWait)
Example:
Configure the /SWEEP END (pin 34) and /PASS FAIL (pin 33)
signals:
CONTrol:HANDler:SWEepend GLOBal
Set the /SWEEP END line to low when all sweeps in all chan-
nels are complete.
CONTrol:HANDler:PASSFail:MODE PASS
Set the default state of the /PASS FAIL line to PASS.
CONTrol:HANDler:PASSFail:SCOPe GLOBal
Set the /PASS FAIL line when all sweeps in all channels are
complete.
CONTrol:HANDler:PASSFail:LOGic POSitive
Set the /PASS FAIL line to positive logic.
CONTrol:HANDler:PASSFail:POLicy ALLTests
Return pass only if all tests pass.
CONTrol:HANDler:PASSfail:POLicy
<Policy>
Specifies how the global pass/fail status (
CONTrol:HANDler:PASSfail:STATus?
Parameters:
<Policy>
ALLTests | ALLMeas
ALLTests
–
the status is PASS if all limit checks in all measure-
ments (traces) pass.
ALLMeas
–
the status is PASS if a limit check is defined for all
measurements (traces) and all limit checks pass. It is FAIL if one
or more traces have no associated limit check, or if at least one
limit check fails.
Note:
*RST
or "Preset" do not change the configuration of the
to restore
default values.
Example:
See
CONTrol:HANDler:PASSfail:SCOPe
<Scope>
Specifies the "sweep end" condition that will cause the /PASS FAIL line (pin 33) to
report the status of the global limit check.
Note:
This setting is not valid if the pass/fail mode is NOWait (
NOWait
).
SCPI command reference