Command reference
R&S
®
ZNB/ZNBT
1033
User Manual 1173.9163.02 ─ 62
<group>
Port group (DUT) number.
If multiple port groups are configured (see
ON
, then each port group can
have its own de-/embedding models.
Parameters:
<Capacitance>
Capacitance C for ground loop embedding.
Range:
-1mF to 1 mF.
Increment: 1 fF (1E-15 F)
*RST:
1 pF (1E-12 F)
Default unit: F
Parameters for setting and query:
<CircuitModel>
SC | SG
Possible circuit models (character data); see
Example:
*RST; :CALC:TRAN:VNET:GLO:EMB:PAR:C? SC
Query the default capacitance for ground loop embedding. The
response is 1E-012 (1 pF).
CALC:TRAN:VNET:GLO:EMB:PAR:C SC, 2.2E-12
Increase the capacitance to 2.2 pF.
Manual operation:
See
CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding<group>:
PARameters:G
<CircuitModel>, <Conductance>
CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding<group>:
PARameters:G?
<CircuitModel>
Specifies the conductance value G in the different circuit models for ground loop
embedding.
Suffix:
<Ch>
.
Channel number.
<group>
Port group (DUT) number.
If multiple port groups are configured (see
ON
, then each port group can
have its own de-/embedding models.
Parameters:
<Conductance>
Conductance G for the specified circuit model.
Range:
-1kS to 1 kS.
Increment: 1 pS (1E-12 S)
*RST:
0 S
Default unit: Siemens (SI unit symbol: S)
Parameters for setting and query:
<CircuitModel>
SG
Possible circuit models (character data); see
SCPI command reference