Operating Manual 1302.6163.12 - 03
6.161
R&S ESU
Remote Control – Description of Commands
SENSe Subsystem
[SENSe<1|2>:]CORRection:TRANsducer:GENerate
<name>
This command generates a transducer factor <name> using normalized trace data. The function is
only available when normalization is switched on.
Parameter:
<name>::= Name of the transducer factors as string data with up to 8 characters.
Example:
"CORR:TRAN:GEN 'FACTOR1'"
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]CORRection:TRANsducer:SELect
<name>
This command selects the transducer factor designated by <name>. If <name> does not exist yet, a
new transducer factor is created.
Parameter:
<name>::= Name of the transducer factor in string data form with a maximum of 8
characters.
Example:
"CORR:TRAN:SEL 'FACTOR1'"
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
R, A
[SENSe<1|2>:]CORRection:TRANsducer:UNIT
<string>
This command specifies the unit for the selected transducer factor. Prior to this command, the
command
SENS:CORR:TRAN:SEL
must be sent.
Parameter:
<string>::= 'DB' | 'DBM' | 'DBMV' | 'DBUV' | 'DBUV/M' | 'DBUA' | 'DBUA/M' | 'DBPW'
| 'DBPT'
Example:
"CORR:TRAN:SEL 'FACTOR1'"
"CORR:TRAN:UNIT 'DBUV'"
Characteristics:
*RST value: 'DB'
SCPI: device-specific
Mode:
R, A
[SENSe<1|2>:]CORRection:TRANsducer:SCALing
LINear | LOGarithmic
This command defines whether the frequency scaling of the transducer factor is linear or logarithmic.
Prior to this command, the command
SENS:CORR:TRAN:SEL
must be sent.
Example:
"CORR:TRAN:SEL 'FACTOR1'"
"CORR:TRAN:SCAL LOG"
Characteristics:
*RST value: LINear
SCPI: device-specific
This command must be sent prior to the subsequent commands for modifying/acti-
vating transducer factors.
Содержание 1129.7246.03
Страница 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Страница 25: ......
Страница 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Страница 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Страница 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Страница 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Страница 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Страница 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Страница 821: ...R S ESU Remote Control Programming Examples More Complex Programming Examples 7 32 Operating Manual 1302 6163 12 03 ...
Страница 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Страница 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Страница 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...