Operating Manual 1302.6163.12 - 03
6.127
R&S ESU
Remote Control – Description of Commands
INITiate Subsystem
This command is an event and therefore has no *RST value and no query.
INITiate<1|2>:DISPlay
ON | OFF
This command configures the behavior of the display during a single sweep.
The numeric suffix of INITiate is irrelevant with this command.
Parameter:
OFF: the display is switched off during the measurement,
ON: the display is switched on during the measurement.
Example:
"INIT2:CONT OFF"
'Switches to single-sweep mode
"INIT2:DISP OFF"
'Sets the display behavior to
OFF
"INIT2;*WAI"
'Starts the measurement with display switched off.
Characteristics:
*RST value: ON
SCPI: device-specific
Mode:
A
INITiate<1|2>:ESPectrum
This command starts a Spectrum Emission Mask measurement. The measurement can be stopped in
continuous sweep mode with
.
Example:
"SENS:SWE:MODE ESP"
' activates the SEM measurement
"INIT:CONT OFF"
' Set to Single Sweep
"INIT:ESP"
' Starts a SEM measurement
Characteristics:
*RST value:-
SCPI: device-specific
Mode:
A
INITiate<1|2>:FMEasurement
In receiver mode, this command starts the final measurement based on the peak list.
Example:
"INIT2:FME"
'Starts the final measurement
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
R
This command is an event, so it has neither an *RST value nor a query function.
Содержание 1129.7246.03
Страница 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Страница 25: ......
Страница 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Страница 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Страница 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Страница 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Страница 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Страница 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Страница 821: ...R S ESU Remote Control Programming Examples More Complex Programming Examples 7 32 Operating Manual 1302 6163 12 03 ...
Страница 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Страница 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Страница 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...