Operating Manual 1302.6163.12 - 03
6.95
R&S ESU
Remote Control – Description of Commands
CALCulate Subsystem
6.5.8
CALCulate:THReshold Subsystem
The CALCulate:THReshold subsystem controls the threshold value for the maximum/minimum search of
markers. The measurement windows are selected by CALCulate 1 (screen A) or 2 (screen B).
CALCulate<1|2>:DLINe<1|2>
MINimum .. MAXimum (depending on current unit)
This command defines the position of Display Line 1 or 2. These lines enable the user to mark any
levels in the diagram. The unit depends on the setting made with
CALC:UNIT
.
Example:
"CALC:DLIN -20dBm"
Characteristics:
*RST value: - (STATe to OFF)
SCPI: device-specific
Mode:
R, A
CALCulate<1|2>:DLINe<1|2>:STATe
ON | OFF
This command switches Display Line 1 or 2 (level lines) on or off.
Example:
"CALC:DLIN2:STAT OFF"
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
R, A
CALCulate<1|2>:THReshold
MINimum to MAXimum (depending on current unit)
This command defines the threshold value for the maximum/minimum search of markers with marker
search functions MAX PEAK, NEXT PEAK, etc. in the selected measurement window. The associated
display line is automatically switched on.
Example:
"CALC:THR -82DBM"
'Sets the threshold value for screen A to -82 dBm.
Characteristics:
*RST value: - (STATe to OFF)
SCPI: device-specific
Mode:
R, A
CALCulate<1|2>:THReshold:STATe
ON | OFF
This command switches on or off the threshold line in the selected measurement window. The unit
depends on the setting performed with
CALC:UNIT
.
Example:
"CALC2:THR:STAT ON"
'Switches on the threshold line in screen B.
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
R, A
Содержание 1129.7246.03
Страница 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Страница 25: ......
Страница 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Страница 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Страница 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Страница 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Страница 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Страница 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Страница 821: ...R S ESU Remote Control Programming Examples More Complex Programming Examples 7 32 Operating Manual 1302 6163 12 03 ...
Страница 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Страница 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Страница 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...