R&S ESU
Remote Control – Description of Commands
CALCulate Subsystem
6.66
Operating Manual 1302.6163.12 - 03
CALCulate<1|2>:MARKer<1...4>:FUNCtion:POWer:RESult?
ACPower | CPOWer | MCACpower |
OBANdwidth | OBWidth | CN | CN0
This command queries the result of the power measurement performed in the selected window.
If necessary, the measurement is switched on prior to the query.
The channel spacings and channel bandwidths are configured in the
SENSe:POWer:ACHannel
subsystem.
To obtain a valid result, a complete sweep with synchronization to the end of the sweep must be
performed before a query is output. Synchronization is possible only in the single-sweep mode.
Parameters:
ACPower:
adjacent channel power measurement
Results are output in the following sequence, separated by commas:
1. Power of transmission channel
2. Power of lower adjacent channel
3. Power of upper adjacent channel
4. Power of lower alternate channel 1
5. Power of upper alternate channel 1
6. Power of lower alternate channel 2
7. Power of upper alternate channel 2
The number of measured values returned depends on the number of adjacent/
alternate channels selected with
SENSe:POWer:ACHannel:ACPairs
.
With logarithmic scaling (RANGE LOG), the power is output in the currently
selected level unit; with linear scaling (RANGE LIN dB or LIN %), the power is
output in W. If
SENSe:POWer:ACHannel:MODE
REL
is selected, the adjacent/
alternate-channel power is output in dB.
AOBandwidth |
AOBWidth
Measurement of occupied bandwidth, all results. The results include the left and
right frequency/level information
Results are output in the following sequence, separated by commas:
1. Occupied bandwidth in Hz
2. T1 marker position in Hz (left marker)
3. T1 marker level
4. T2 marker position in Hz (right marker)
5. T2 marker level
Note:
The Occupied Bandwidth is marker position T2 - T1
CPOWer:
Channel power measurement
With logarithmic scaling (RANGE LOG), the channel power is output in the cur-
rently selected level unit; with linear scaling (RANGE LIN dB or LIN %), the chan-
nel power is output in W.
MCACpower: Channel/adjacent
channel
power measurement with several carrier signals
Results are output in the following sequence, separated by commas:
1. Power of carrier signal 1
2. Power of carrier signal 2
Содержание 1129.7246.03
Страница 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Страница 25: ......
Страница 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Страница 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Страница 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Страница 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Страница 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Страница 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Страница 821: ...R S ESU Remote Control Programming Examples More Complex Programming Examples 7 32 Operating Manual 1302 6163 12 03 ...
Страница 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Страница 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Страница 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...