
Operating Manual 1302.6163.12 - 03
6.47
R&S ESU
Remote Control – Description of Commands
CALCulate Subsystem
CALCulate<1|2>:MARKer<1...4>:Y:PERCent
0 to100%
This command positions the selected marker in the selected window to the given probability. If marker
2, 3 or 4 is selected and used as a delta marker, it is switched to marker mode.
Example:
"CALC1:MARK:Y:PERC 95PCT"
'Positions marker 1 in screen A to a 'probability of 95%.
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
A
CALCulate<1|2>:MARKer<1...4>:MAXimum[:PEAK]
This command positions the specified marker to the current maximum value of the corresponding
trace. The corresponding marker is activated first or switched to the marker mode.
Example:
"CALC:MARK2:MAX"
'Positions marker 2 in screen A to the maximum value of 'the trace.
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
R, A
This command is an event and therefore has no *RST value and no query.
CALCulate<1|2>:MARKer<1...4>:MAXimum:NEXT
This command positions the specified marker to the next smaller maximum value of the corresponding
trace.
Example:
"CALC:MARK2:MAX:NEXT"
'Positions marker 2 in screen A to the next 'lower maximum value.
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
R, A
This command is an event and therefore has no *RST value and no query.
The command is only available with the CCDF measurement switched on.
The associated level value can be determined with the
CALC:MARK:X?
command.
If no maximum value is found on the trace (level spacing to adjacent values < peak
excursion), an execution error (error code: -200) is produced.
If no next smaller maximum value is found on the trace (level spacing to adjacent
values < peak excursion), an execution error (error code: -200) is produced.
Содержание 1129.7246.03
Страница 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Страница 25: ......
Страница 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Страница 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Страница 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Страница 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Страница 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Страница 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Страница 821: ...R S ESU Remote Control Programming Examples More Complex Programming Examples 7 32 Operating Manual 1302 6163 12 03 ...
Страница 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Страница 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Страница 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...