R&S ESU
Remote Control – Basics
Status Reporting System
5.26
Operating Manual 1302.6163.12 - 03
5.7.3.7
STATus:QUEStionable:FREQuency Register
This register comprises information about the reference and local oscillator.
It can be queried with commands
STATus:QUEStionable:FREQuency:CONDi-
tion?
and
STATus:QUEStionable:FREQuency[:EVENt]?
.
10
ALT1 UPPer FAIL (Screen B)
This bit is set if in screen B the limit is exceeded in the upper 1st alternate channel.
11
ALT1 LOWer FAIL (Screen B)
This bit is set if in screen B the limit is exceeded in the lower 1st alternate channel.
12
ALT2 UPPer FAIL (Screen B)
This bit is set if in screen B the limit is exceeded in the upper 2nd alternate channel.
13
ALT2 LOWer FAIL (Screen B)
This bit is set if in screen B the limit is exceeded in the lower 2nd alternate channel.
14
This bit is not used.
15
This bit is always set to 0.
Bit
No.
Meaning
Bit
No.
Meaning
0
OVEN COLD
This bit is set if the reference oscillator has not yet attained its operating temperature. 'OCXO'
will then be displayed.
1
LO UNLocked
(Screen A)
This bit is set if the local oscillator no longer locks. 'LOUNL will then be displayed.
2 to 8
not used
9
LO UNLocked
(Screen B)
This bit is set if the local oscillator no longer locks.' LOUNL' will then be displayed.
10 to
14
not used
15
This bit is always 0.
Содержание 1129.7246.03
Страница 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Страница 25: ......
Страница 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Страница 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Страница 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Страница 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Страница 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Страница 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Страница 821: ...R S ESU Remote Control Programming Examples More Complex Programming Examples 7 32 Operating Manual 1302 6163 12 03 ...
Страница 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Страница 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Страница 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...