R&S ESU
Remote Control – Basics
Getting Started
5.4
Operating Manual 1302.6163.12 - 03
5.3 Getting Started
The short and simple operating sequence provided below enables you to quickly put
the instrument into operation and set its basic functions. As a prerequisite, the GPIB
address, which is factory-set to 20, must remain unchanged.
1. Connect instrument and controller using GPIB cable.
2. Write and start the following program on the controller:
CALL IBFIND("DEV1", analyzer%)
'Open port to the instrument
CALL IBPAD(analyzer%, 20)
'Inform controller about instrument address
CALL IBWRT(analyzer%, '*RST;*CLS')
'Reset instrument
CALL IBWRT(analyzer%, 'FREQ:CENT 20MHz')
'Set center frequency to 20 MHz
CALL IBWRT(analyzer%, 'BAND 9 kHz')
'Set resolution bandwidth to 9 kHz
CALL IBWRT(analyzer%, 'DET:REC POS')
'Switch on peak detector
The instrument now measures at 20 MHz with 9 kHz RBW the peak level.
3. To return to manual operation, press the
LOCAL
key at the front panel.
5.4 Starting Remote Control
On power-on, the instrument is always in the manual operating state ("LOCAL"
state) and can be operated via the front panel.
It is switched to remote control ("REMOTE" state)
During remote control, operation via the front panel is disabled. The instrument
remains in the remote state until it is reset to the manual state via the front panel or
via remote control interfaces. Switching from manual operation to remote control
and vice versa does not affect the remaining instrument settings.
5.4.1
Display Contents during Remote Control
During remote control, only the LOCAL softkey appears, with which it is possible to
return to manual operation.
In addition, the display of diagrams and results can be blanked out with the com-
mand
"SYSTem:DISPlay:UPDate OFF"
(default in remote control) to obtain opti-
mum performance during remote control.
GPIB
as soon as it receives an addressed command from a controller.
if it is controlled in a network (RSIB interface), as soon as it receives a command
from a controller.
RS-232
as soon as it receives the command "
@
REM" from a controller.
Содержание 1129.7246.03
Страница 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Страница 25: ......
Страница 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Страница 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Страница 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Страница 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Страница 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Страница 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Страница 821: ...R S ESU Remote Control Programming Examples More Complex Programming Examples 7 32 Operating Manual 1302 6163 12 03 ...
Страница 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Страница 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Страница 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...