R&S ESU
Instrument Functions
Basic Settings
4.272
Operating Manual 1302.6163.12 - 03
If transducers are active the function is automatically set to AUTO to obtain the best
dynamic performance.
Remote command:
CORR:TRAN:SEL 'FACTOR1'
CORR:TRAN:ADJ:RLEV ON | OFF
PAGE UP
The
PAGE UP
softkey is used to scroll through large tables that cannot completely
be displayed on the screen.
PAGE DOWN
The
PAGE DOWN
softkey is used to scroll through large tables that cannot com-
pletely be displayed on the screen.
Entry and Editing of Transducer Factors
A transducer factor is characterized by the following:
• Reference values with frequency and factor (
Values
)
• Unit of the factor (
Unit
) and
• Name (
Name
) to distinguish the various factors.
During entry the R&S ESU checks the transducer factor for compliance with specific
rules that must be met to ensure correct operation.
• The frequencies for the reference values must always be entered in ascending
order. Otherwise the entry will not be accepted and the following message is
displayed:
WRONG FREQUENCY SEQUENCE !
• The frequencies entered may exceed the frequency range of the R&S ESU since
only the set frequency range is taken into account for measurements. The
minimum frequency of a reference value is 0 Hz, the maximum frequency
200 GHz.
• The value range for the transducer factor is ±200 dB. If the minimum or maximum
value is exceeded, the R&S ESU outputs the following message:
out of range
• Gain has to be entered as a negative value, and attenuation as a positive value.
NEW / EDIT
The submenu differs for transducer factors and tranducer sets. The submenu for
transducer sets is described in section
“Entry and Editing of Transducer Sets” on
. In the following, the submenu for transducer factors is described.
The
NEW
and
EDIT
softkeys give access to the submenu for editing and generating
transducer factors.
The softkeys in the
UNIT
submenu of the
AMPT
key cannot be operated if the
transducer is on.
Содержание 1129.7246.03
Страница 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Страница 25: ......
Страница 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Страница 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Страница 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Страница 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Страница 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Страница 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Страница 821: ...R S ESU Remote Control Programming Examples More Complex Programming Examples 7 32 Operating Manual 1302 6163 12 03 ...
Страница 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Страница 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Страница 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...