Operating Manual 1302.6163.12 - 03
4.205
R&S ESU
Instrument Functions
Analyzer Mode
MEAN PWR
POSITION
Defines the relative X position of the mean power value for the CCDF measurement.
The default position is 0% (left corner of the grid). This softkey is available if the
CCDF measurement is switched on.
Remote command:
CALC:STAT:SCAL:X:MPOS <numeric_value>
ADJUST
SETTINGS
“ADJUST SETTINGS” on page 4.204
CONT MEAS
The
CONT MEAS
softkey starts collecting a new sequence of sample data and cal-
culating the APD or CCDF curve depending on the selected measurement. The next
measurement is started automatically as soon as the indicated number of samples
has been reached.
Remote command:
INIT:CONT ON;
INIT:IMM
SINGLE MEAS
The
SINGLE MEAS
softkey starts collecting a new sequence of sample data and
calculating the APD or CCDF curve depending on the selected measurement. At the
beginning of the measurement previously obtained measurement results are dis-
carded.
Remote command:
INIT:CONT OFF;
INIT:IMM
Hint for usage of the marker functions with measurement of signal
statistics
With the signal statistic measurement level always is displayed on x-axis. Y-axis
always is a normalized value between 0 and 1. In contrary to use of marker in fre-
quency or time domain marker is input in level values and the output is in percent-
age values.
Example
Measurement of CCDF of a IS95 BTS signal, level 0 dBm, frequency 800 MHz
[
PRESET
]
Switch on preset settings.
[
FREQ
:
CENTER
: 800
MHz]
Set center frequency to 800 MHz.
[
AMPT
: 10 dBm]
Set reference level to 10 dBm.
[
BW
: 3 MHz]
Set resolution bandwidth to 3 MHz (resolution bandwidth
shall be wider then signal bandwidth (1.25 MHz) in order
to have the complete signal within the resolution band-
width).
[
MEAS
]
Call the menu for measurement functions.
[
SIGNAL STATISTIC
]
Call the menu for signal statistics measurement.
[
CCDF ON /OFF
]
Switch on measurement of the complementary cumulative
distribution function. The R&S ESU switches to zero span
mode. The power of the signal and the CCDF is calculated
for the number of samples selected. With the CCDF func-
tion sample detector and video bandwidth are set auto-
matically.
Содержание 1129.7246.03
Страница 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Страница 25: ......
Страница 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Страница 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Страница 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Страница 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Страница 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Страница 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Страница 821: ...R S ESU Remote Control Programming Examples More Complex Programming Examples 7 32 Operating Manual 1302 6163 12 03 ...
Страница 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Страница 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Страница 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...