Operating Manual 1302.6163.12 - 03
4.173
R&S ESU
Instrument Functions
Analyzer Mode
STOP LIMIT
The
STOP LIMIT
softkey activates the entry of the upper limit of the evaluation
range.
Remote command:
CALC:MARK:X:SLIM:RIGH <value>
SET
REFERENCE
The
SET REFERENCE
softkey sets the power values currently measured as refer-
ence values for the calculation of the mean value (
MEAN)
and the rms value (
RMS)
.
The reference values are used to perform relative measurements.
If the calculation of the mean value (
MEAN)
and rms value (
RMS)
is not switched
on, 0 dBm is used as a reference value.
If the average value (
AVERAGE)
or maximum value (
MAX HOLD
) is calculated over
several sweeps, the current value is the measured value summed up at the actual
time.
Remote command:
CALC:MARK:FUNC:SUMM:REF:AUTO ONCE
POWER
ABS | REL
The
POWER ABS/REL
softkey selects the absolute power measurement (default
setting) or relative power measurement. The reference value for the relative power
is defined by
SET REFERENCE
.
The value 0 dBm is used if the reference value is not defined.
Remote command:
CALC:MARK:FUNC:SUMM:MODE ABS
MAX HOLD
ON | OFF
The
MAX HOLD ON/OFF
softkey switches the display of the maximum peak
obtained from measurements at successive sweeps on and off.
The displayed maximum peak is only updated at the end of a sweep if a higher
value has occurred.
The maximum value can be reset by switching the
MAX HOLD ON / OFF
softkey off
and on again.
Remote command:
CALC:MARK:FUNC:SUMM:PHOL ON
CALC:MARK:FUNC:SUMM:PPE:PHOL:RES?
CALC:MARK:FUNC:SUMM:RMS:PHOL:RES?
CALC:MARK:FUNC:SUMM:MEAN:PHOL:RES?
CALC:MARK:FUNC:SUMM:SDEV:PHOL:RES?
AVERAGE
ON | OFF
The
AVERAGE ON/OFF
softkey switches averaging over successive sweep mea-
surements on and off.
The measured values can be reset by switching the
AVERAGE ON / OFF
softkey off
and on again.
Remote command:
CALC:MARK:FUNC:SUMM:AVER ON
CALC:MARK:FUNC:SUMM:PPE:AVER:RES?
CALC:MARK:FUNC:SUMM:RMS:AVER:RES?
CALC:MARK:FUNC:SUMM:MEAN:AVER:RES?
CALC:MARK:FUNC:SUMM:SDEV:AVER:RES?
Содержание 1129.7246.03
Страница 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Страница 25: ......
Страница 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Страница 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Страница 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Страница 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Страница 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Страница 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Страница 821: ...R S ESU Remote Control Programming Examples More Complex Programming Examples 7 32 Operating Manual 1302 6163 12 03 ...
Страница 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Страница 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Страница 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...