This low average value is usually corrected in noise power measurements by a 2.5
dB factor. Therefore the R&S ESU offers the selection of linear averaging. The trace
data are delogarithmized prior to averaging, then averaged and logarithmized again
for display on the screen. The average value is always correctly displayed
irrespective of the signal characteristic.
COPY TRACE
The
COPY TRACE
softkey copies the screen contents of the current trace into
another trace memory. The desired memory is selected by entering the number 1, 2
or 3.
Upon copying, the contents of the selected memory are overwritten and the new
contents displayed in view mode.
Remote command:
TRAC:COPY TRACE1,TRACE2
Содержание 1129.7246.03
Страница 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Страница 25: ......
Страница 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Страница 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Страница 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Страница 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Страница 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Страница 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Страница 821: ...R S ESU Remote Control Programming Examples More Complex Programming Examples 7 32 Operating Manual 1302 6163 12 03 ...
Страница 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Страница 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Страница 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...