MTR302/A - R302F/A - R302L/A DHHS
MANUALE ISTRUZIONI / INSTRUCTION MANUAL
Modello/Model R302/A
Compliance Verification
8
R302/A
- Compliance V
erification
Ralco
srl
Electromagnetic Immunity for All Equipment and Systems
C
OLLIMATOR
R302/A
IS
INTENDED
FOR
USE
IN
THE
ELECTROMAGNETIC
ENVIRONMENMT
SPECIFIED
BELOW
. T
HE
CUSTOMER
OR
THE
USER
SHOULD
ASSURE
THAT
IT
IS
USED
IN
SUCH
AN
ENVIRONMENT
.
Immunity
Test
IEC 60601 Test
Level
Complian
ceLevel
Electromagnetic Environment
Electrostatic
discharge
(ESD)
IEC 61000-4-2
6 kV contact
8 kV air
EN 60601-
1-2
test level
Hospital
Electronical
fast transient/
burst
IEC 61000-4-4
2 kV for power
supply lines
1 kV for input/
output lines >3m
EN 60601-
1-2
test level
Hospital
Surge
IEC 61000-4-5
1 kV differential
mode
2 kV common
mode
EN 60601-
1-2
test level
Hospital
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-
11
0% U
n
for 0.5
cycles.
40% U
n
for 5
cycles
70% U
n
for 25
cycles
0% U
n
for 5 s
EN 60601-
1-2
test level
Hospital
Power
frequency (50/
60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
EN 60601-
1-2
test level
Hospital