Model R 221 ACS - R 221 ACS DHHS
MTR 221 ACS - MTR 221 ACS DHHS
MANUALE ISTRUZIONI / INSTRUCTION MANUAL
C
AN
B
US
M
ESSAGES
61
MTR
2
21
AC
S
-
MTR 22
1
ACS DHHS - Can Bus Me
ssages
C
OLLIMATOR
C
ONFIGURATION
Rx_Status
0x7F5
8
Status Message retrieved from the Collimator.
Rx_Status
0x7F7
8
Status Message retrieved from the Collimator.
Rx_Acknowledge
0x7F8
8
Acknowledge Message retrieved from the Collimator.
Rx_Status
0x7F9
8
Status Message retrieved from the Collimator.
Rx_Status
0x7FC
8
Status Message retrieved from the Collimator.
Tx_DisplayConfig
0x7C0
8
Message sent to the collimator in order to set its actual LCD Display Visualization Type.
Tx_DisplayConfig
0x7C1
8
Message sent to the collimator in order to set its Welcome Message (Valid for LCD Free Mode Visualization Type only).
Rx_Status
0x7D0
2
Status Message retrieved from the Collimator.
Rx_Status
0x7D0
8
Status Message retrieved from the Collimator.
Rx_Status
0x100
2
Status Message retrieved from the Collimator.
Rx_Status
0x100
8
Status Message retrieved from the Collimator.
Type
ID
DLC RTR
D0
D1
D2
D3
D4
D5
D6
D7
Description
Fro
m
FW
Ver
sion
Tx_Con
fig
0x600
8
0
0x00
0x00
0x00
0x00
0x00
0x00
0x00
0x00
Requires to the Collimator to visualize on the LCD display the
string "RS232COMM".
1.00
Tx_Con
fig
0x600
8
0
0x01
MSB_SerialNr
CSB_SerialNr
CSB_Seri
alNr
CSB_Seri
alNr
LSB_Seri
alNr
0x00
0x00
Store into the GC007 Flash Memory the X-Ray Collimator serial
number with the provided value.
1.00
Tx_Con
fig
0x600
8
0
0x02
Coll_Language
0x00
0x00
0x00
0x00
0x00
0x00
Configure the X-Ray Collimator language displayed on the LCD
screen with the provided setting value.
Coll_Language : 0x01 = Italian.
Coll_Language : 0x02 = English. (default)
1.00
Tx_Con
fig
0x600
8
0
0x03
Meas_Unit
0x00
0x00
0x00
0x00
0x00
0x00
Configure the X-Ray Collimator measurement unit displayed on
the LCD screen with the provided setting value.
Meas_Unit : 0x01 = Metric System. (default)
Meas_Unit : 0x02 = Imperial System.
1.00
Tx_Con
fig
0x600
8
0
0x04
Show_SID
0x00
0x00
0x00
0x00
0x00
0x00
Configure the X-Ray Collimator SID visualization displayed on
the LCD screen with the provided setting value.
Show_SID : 0x00 = The LCD screen does not displays the SID
value.
Show_SID : 0x01 = The LCD screen displays the SID value.
(default)