Configuration of the collimator via serial interface is performed in Ralco with updated software;
the customer will be required to load the software as described in the preceding paragraph.
P
OTENTIOMETER
P
OSITIONING
A
LARMS
The alarm may be enabled in message 0x100 (0x7 D0) through 0x600 entering control 0x33 in
the configuration phase.
This message includes the control bytes for potentiometer positioning alarms.
When the collimator activates a movement alarm, the system will read the collimator alarm
state through message 0x100 and the number of alarms activated from start-up.
The red LED, Exposure Hold, on the collimator front panel flashes to indicate an alarm state.
Alarm State Reset
Turn the collimator key from automatic to manual. This will reset the shutters and Iris if
assembled.
Possible Root Causes of the problem:
• The Bootloader is not installed on the GC007 PCB.
• The X-Ray Collimator is not powered on.
• The GC007 PCB is defective or broken.
• There are more than one COM devices active on the PC, and the Loader_2.1 application
program is trying to upload the firmware on the wrong one.