MTR 221 ACS - MTR 221 ACS DHHS
MANUALE ISTRUZIONI / INSTRUCTION MANUAL
S
YSTEM
S
TART
-U
P
57
Model R 221 ACS - R 221 ACS DHHS
MTR 221
ACS
-
MTR 221
ACS
DHHS
- Sy
stem Star
t-Up
HW Setup
connect the USB Serial Converter adapter to
the PC and to the Serial Port placed into the
GC007 PCB (visible by unmounting the XRay
Collimator front panel).
Connect the X-Ray Collimator to the power
supply but keep it still powered OFF.
Config
Into the main folder of the Loader_2.1
Application program there is the file
seriale.txt.
Such file is used by the program to set the
following configurations:
1. COM Port Number : it’s the number of the
COM port to use. Such number shall
match exactly the number of the COM port
2. detected by the Operating System.
For instance, if Windows configures the USB Serial converter as COM5 (see image below).
The file seriale.txt shall be modified as follows:
COM =
5
BAUD_RATE = 115200
3. COM Baud Rate : it’s the baud rate used for the file transferring procedure. (that is equal
to 115200kHz for the X-Ray Collimators).
Copy
Copy the .COD file into the same folder where are stored both Loader_2.1.exe and seriale.txt
files.
Such file contains the Firmware to upgrade into the X-Ray Collimator.
Important note:
Use only.COD firmware files provided and approved by RALCO S.r.l. Any other file different
from these ones can be responsible of a malfunctioning of the collimator.