MTR 221 ACS - MTR 221 ACS DHHS
MANUALE ISTRUZIONI / INSTRUCTION MANUAL
O
PTIONAL
I
TEMS
210
Model R 221 ACS - R 221 ACS DHHS
MT
R
2
21
ACS
-
MTR
2
21
AC
S DHH
S - Optio
nal Ite
m
s
EMC C
OMPLIANCE
Electromagnetic Emissions
Electromagnetic Immunity for All Equipment and Systems
The R 221 ACS - R 221 ACS DHHS colimator is suitable for use in the specified electromagnetic environment. The purchaser
or user of the R 221 ACS - R 221 ACS DHHS should assure that it is used in an electromagnetic environment as described
below:
Emissions Test
Compliance
Electromagnetic Environment- Guidance
RF emissions
CISPR 11
Group 1
The Collimator R 221 ACS - R 221 ACS DHHS needs special
precautions regarding EMC and needs to be installed and put
into service according to the EMC report.
Portable and mobile RF communications equipment can affect
the collimator R 221 ACS - R 221 ACS DHHS.
RF emissions
CISPR 11
Class [A]
This R 221 ACS - R 221 ACS DHHS is suitable for use in all
establishments other than domestic and those directly connected
to the low voltage power supply network which supplies buildings
used for domestic purposes.
Harmonic distortion
IEC 61000-3-2
[Not applicable]
Limitation of Voltage Fluctuation
and Flicker
IEC 61000-3-3
[Not applicable]
This R 221 ACS - R 221 ACS DHHS is not suitable to be
connected to other equipments.
Collimator R 221 ACS - R 221 ACS DHHS is intended for use in the electromagnetic environmenmt specified below. The
customer or the user should assure that it is used in such an environment.
Immunity
Test
IEC 60601 Test
Level
Complian
ce Level
Electromagnetic Environment
Electrostatic
discharge (ESD)
IEC 61000-4-2
8 kV contact
15 kV air
EN 60601-1-2
test level
Hospital