MTR 221 ACS - MTR 221 ACS DHHS
MANUALE ISTRUZIONI / INSTRUCTION MANUAL
A
DJUSTMENTS
105
Model R 221 ACS - R 221 ACS DHHS
MTR 221
ACS
-
MTR 2
21
AC
S DHH
S - Adjustments
ADJUSTMENTS
C
ROSSHAIR
A
DJUSTMENT
• Activate the light field.
• Adjust the light field to a narrow line for each pair of shutters
by turning the two knobs alternately.
• Check that the project cross line is exact halfway between
the edges of the shutters, see
Fig. Cross Line
.
• If adjustment is required, remove the cover from the sides
and bottom of the collimator, see Chapter -
C
OVER
R
EMOVAL
.
• Loosen the four screws securing the plastic panel and adjust
the cross lines to coincide with the light lines, see Chapter -
C
OVER
R
EMOVAL
.
• Tighten the screws.
L
ASER
A
DJUSTMENT
• Remove part of the cover to access
the point of adjustment, see Chapter-
C
OVER
R
EMOVAL
.
• The line is to fall on a perpendicular
cross-line on the plastic anti-dust
panel near the collimator controls,
see
Fig.
Laser Line.
• Adjust the position of the line by rotat-
ing the laser or moving the base of
the laser system.
• To rotate the laser system,
loosen the Allen screw
A
, see
Fig. Laser Adjustment.
• Tighten the Allen screw when
the laser beam falls on or is
parallel to the bisector line
drawn on the anti-dust panel.
• Tilt the laser system by loosening the
two
B
screws holding the laser system base to the beam limiting device front plate.
W
ARNINGS
T
HIS
IS
PROVIDED
AS
AN
AID
TO
THE
END
USER
. R
ALCO
IS
NOT
LIABLE
FOR
ANY
DAMAGES
RESULTING
FROM
THE
ALTERING
OF
PRESET
FACTORY
CONFIGURATIONS
.
A
DJUSTMENTS
HAVE
BEEN
PRESET
AT
THE
FACTORY
PRIOR
TO
SHIPMENT
OF
THE
COLLIMATOR
.
SHOULD
ANY
PROBLEM
ARISE
REQUIRING
THE
NEED
TO
RECALIBRATE
,
PLEASE
CONTACT
RALCO
FOR
ASSISTANCE
BEFORE
PROCEEDING
WITH
THE
ADJUSTMENT
.
Fig. Cross Line