Electromagnetic Immunity for All Equipment and Systems
Collimator R 221 ACS - R 221 ACS DHHS is intended for use in the electromagnetic environmenmt specified below. The
customer or the user should assure that it is used in such an environment.
Immunity
Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic Environment
Electrostatic
discharge (ESD)
IEC 61000-4-2
8 kV contact
2/4/8/15 kV air
EN 60601-1-2
test level
Floors should be wood. concrete or ceramic tile. If floors
are covered with synthetic material. the relative humidity
should be at least 30%.
Radiated
electromagnetic
field
IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
IEC 60601-1-2
Test level
Portable and mobile RF communications equipment
should be used no closer to any part of the
collimator
R
221 ACS - R 221 ACS DHHS. including cables. Mini-
mum distance 30 cm
Electronical fast
transient/burst
IEC 61000-4-4
2 kV for power
supply lines
1 kV for input/output
lines >3m
EN 60601-1-2
test level
Mains power quality should be that of a typical commer-
cial or hospital environment.
Surge
IEC 61000-4-5
0.5/1 kV differential
mode
0.5/1/2 kV common
mode
EN 60601-1-2
test level
Mains power quality should be that of a typical commer-
cial or hospital environment.
Conducted
disturbances
induced by RF fields
IEC 61000-4-6
3 V
150 kHz to 80 MHz
6V
ISM frequencies
IEC 60601-1-2
Test level
Portable and mobile RF communications equipment
should be used no closer to any part of the
collimator
R
221 ACS - R 221 ACS DHHS. including cables.
Minimum distance 30 cm
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
10 ms – 0% a 0°, 45°,
90°, 135°, 180°. 225°,
270°, 315°
20 ms – 0% a 0°
500 ms – 70% a 0°
5 s – 0%
EN 60601-1-2
test level
Mains power quality should be that of a typical commer-
cial or hospital environment. If the user of the
collimator
R 221 ACS - R 221 ACS DHHS requires continued op-
eration during power mains interruptions. it is recom-
mended that the
collimator
R 221 ACS - R 221 ACS
DHHS
be powered from an uninterruptible power supply
or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
EN 60601-1-2
test level
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial
or hospital environment.
Frequency Range and Level: RF wireless communication EQUIPMENT
Collimator R 221 ACS - R 221 ACS DHHS is intended for use in the electromagnetic environmenmt specified below. The
customer or the user should assure that it is used in such an environment.
Test
Frequency
(MHz)
Modulation
Minimum
IMMUNITY
Level (V/m)
IMMUNITY Level
Applied (V/M)
385
**Pulse Modulation: 18 Hz
27
27
450
*FM + 5 Hz deviation: 1 kHz sine
**Pulse Modulation: 18 Hz
28
28