Transversal Calibration (CROSS)
If the light-field needs calibration, the
mirror needs to be adjusted as follows:
• Remove the part of the cover neces-
sary to access the screws, see Chap-
ter -
C
OVER
R
EMOVAL
.
• Loosen the mirror fixing screw
A
(not
remove) and shift it to adjust the posi-
tion of the mirror, see
Fig. Transver-
sal Calibration (CROSS).
• Once you have regulated the mirror
tighten the screw
A
and remount the
cover, see Chapter -
C
OVER
R
EMOVAL
.
GC-LED-5A
C
ONNECTORS
J2 - S
UPPLY
AND
O
UTPUTS
1. 20-30 V AC or 12-45 V DC
2. 20-30 V AC or 0 V DC
3. +5 V Fan
4. GND Fan
5. +5 V Laser
6. GND Laser
7. + Power-up/Power down external push-button
8. GND Power-up external push-button
J1 - LED O
UTPUT
+ LED 5 A positive output
- GND Led 5A
Fig. Transversal Calibration (CROSS)