ToxiRAE 3 QuickStart Guide
RAE Systems by Honeywell
3775 N. First St.
San Jose, CA 95134-1708 USA
Phone:
408-952-8200
Fax:
408-952-8480
Email:
Web Site:
www.raesystems.com
WARNINGS
Read Before Operating
The User’s Guide must be carefully read by all
individuals who have or will have the responsibility
of using, maintaining, or servicing this product.
The product will perform as designed only if it is
used, maintained, and serviced in accordance with
the manufacturer’s instructions.
CAUTION!
Never operate the monitor when the cover is
removed. Remove monitor cover and battery only
in an area known to be non-hazardous.
User Interface
The ToxiRAE 3’s user interface consists of the
display, LEDs, an alarm buzzer, and two keys,
[MODE] and [Y/+]. The LCD display provides
visual feedback that includes calibration or bump
request, sensor mode, and battery condition.
Turning The ToxiRAE 3 On
Press and hold [MODE]. The display shows “On” and
the sensor type, and then alarms and sensor are tested.
Note:
If the sensor is depleted or does not check out
satisfactorily, an error message of “E01” flashes on
the display. The ToxiRAE 3 should not be used. (Shut
off the ToxiRAE 3 by pressing and holding [MODE]
until you see the countdown and the display shows
“oFF.”).
Once the countdown is complete, a battery icon that
indicates battery condition. “OK” is displayed below
the battery icon, and the reading is also shown (for
example 0 ppm). The ToxiRAE 3 is now in Normal
Mode and is ready for use.
Turning The ToxiRAE 3 Off
Press and hold [MODE]. In 2 seconds, a 5-second
countdown to shutoff begins. You must hold your
finger on the key for the entire shutoff process. If
you remove your finger from the key during the
countdown, the shutoff operation is canceled and
the ToxiRAE 3 continues normal operation.
When you see “oFF,” release your finger from the
[MODE] key. The ToxiRAE 3 is now off.
Low Battery Warnings
•
When the battery’s charge is low, the battery
symbol flashes once per second and “bAt”
flashes once per minute. The buzzer and vibrator
alarm once per minute.
•
When the battery reaches a critically low charge,
the battery symbol flashes once per second and
“bAt” flashes once per minute, plus the buzzer
sounds once per minute.
•
When the battery is dead, the buzzer sounds
once per second for the last minute and then
shuts off. The display shows “oFF.”
Important!
Do not operate the ToxiRAE 3 again
until its battery has been replaced.
WARNING!
To reduce the risk of ignition of hazardous
atmospheres, replace battery only in areas
known to be non-hazardous. Consult User’s
Guide for battery replacement procedure.
Belt clip
LEDs
Alarm
Buzzer
Display
[MODE] key
[Y/+] key
Gas inlet
2.
3.
5.
6.
4.