Platform 2:
Icon Step
Value
Description
1. [CF] Flags
condition
Signal flag 1
– „1”
Checking condition, whether signal flag 1 is set to value
“1” – i.e. checking whether desired part of the process
has been already realized on the weighing platform no.
1. If Yes, then dosing process on the weighing platform
no. 2 shall start.
2. [T] Tare
Tare
Tarring the weighing platform no. 2
3. [TI] Delay
[5s] Open
valve Water
Pending for opening main valve of the container with
“Water”
4. [DA]
Automatic
batching
2 kg [Water]
Automatic dosing of ingredient “Water” to obtain 2 kg
mass in single-step dosing mode – Output 6 controls
the dosing valve (set for Water in Products database)
6. [TI] Delay
[5s] Close
valve Water
Pending for closing the main valve of the container
“Water”
7. [O]
Outputs
Output 12 –
„1”
Output 12 akes high status (“1”) – signal confirming
carrying out the dosing process on the weighing
platform no. 2 is lit up
8. [F] Set
flags
Signal flag 1
– „1”
Signal flag 2
– „1”
Setting the characteristic point of the process, which is
a condition for carrying out part of the process on the
weighing platform no. 1
8. [TI] Delay
[10s] Dosing
completed
Displaying a text message on the terminal confirming
completion of dosing process
An instance of mixed dosing (manual and automatic) has been extended by
signalling its corresponding phases to present the possibilities of the working
mode Formulation.
27.7. Reporting from completed formula making processes
On completing each formula making process, the terminal automatically
generates a report on that process.
Notice:
Submenu: “
Devices /
Printer /
Printouts /
Formulation
report printout template” enables optional modifying of the report template
(see ch. 16.2.3 of this user manual).
98
Содержание WLY/KTP Series
Страница 2: ...2 DECEMBER 2011...
Страница 10: ...Y KTP C1 R series main dimensions Y KTP C1 K series main dimensions 5 1 2 Scales of Y KTP D2 series 10...
Страница 150: ...Scale computer cable Scale Kafka printer cable Scale printer CITIZEN EPSON cable Scale I O cable 150...
Страница 175: ...Print the marked report on a conneted printer by pressing Using button view the selected report 175...