23. WORKING MODE – PERCENT WEIGHING
Percent weighing is a working mode enabling you to compare measured load
mass with the reference sample mass. The result is expressed in [
%
].
Reference sample mass can be either determined by weighing or entered
to weighing device memory by an operator.
23.1. Local Settings
To go to local settings enter
<3.4.dEu>
submenu.
3.4.1.Acc
Working mode accessibility
For detailed description read section 19.2.1.
3.4.2.UUt
Operation mode
For detailed description read section 23.1.1.
3.4.3.Snn
Save mode
For detailed description read section 19.2.2.
3.4.4.Int
Time interval
For detailed description read section 19.2.3.
3.4.5.Lo
Lo threshold
For detailed description read section 19.2.4.
23.1.1. Selecting Operation Mode
Parameter allowing you to select method of determination of reference weight
mass.
Procedure:
•
Enter
<3.4.dEu / 3.4.2.UUt>
submenu.
•
Press
key, parameter values are displayed successively one by one:
S_S
Select to set reference sample mass by determining the mass value.
Suu
Select to set reference sample mass by entering the mass value.
•
Set respective value and press
key to confirm, next go to the home
screen.
23.2. Reference Sample Mass Determined by Weighing
•
Enter
<3.4.dEu / 3.4.2.UUt>
submenu, set
<
S_S
>
value.
•
Enter
<dEu>
working mode (Percent weighing), first, text
<LoAd>
is displayed for 1 second, then the weighing window.
•
Load the weighing pan with the reference sample. When the indication
is stable (
pictogram is displayed), press
key to confirm the mass.