11.3. Taring
To determine net weight value, load the weighing pan with a packaging, wait
for a stable indication and press
key. Zero indication and the following
pictograms are displayed:
Net
and
. The weighing device has been tared.
Upon loading, net mass is displayed.
Taring can be carried out repeatedly within the whole weighing range.
Remember not to exceed the maximum capacity, i.e. sum of tare weight value
and load weight value must be lower than the maximum capacity value. Upon
unloading the weighing pan, the sum of tared masses with minus sign
is displayed.
It is impossible to tare zero or negative values. When you
tare zero or negative values, message <Err3> is displayed,
short sound signal is heard.
11.4. Entering Tare Value Manually
•
Press
and
keys combination, tare value edit box is displayed.
•
Enter tare value, to do it press
and
keys:
Press to select digit that is to be edited.
Press to set digit value, 0 - 9 .
•
Press
key to confirm, the scale returns to the weighing mode,
modified tare value with
'–'
sign is displayed.
•
You can enter tare value at any time during the weighing operation.
11.5. Dual Range Devices
N/A in case of single range scales
Switching from weighing with the accuracy of the
I weighing range
to weighing
with the accuracy of the
II weighing range
takes place automatically upon
exceeding Max of the
I weighing range
. Upon switching to weighing with the
accuracy of the
II weighing range
, respective pictogram/marker
is displayed in the top left hand corner.