background image

 

19 

 

11.  TECHNICAL SPECIFICATIONS 

 

For  technical  specifications  of  respective  scales  go  to  RADWAG  website 

www.radwag.pl

 

 
 

12.  TROUBLESHOOTING 

 

Problem 

Cause 

Solution 

Scale start-up fail. 

Power supply disconnected. 

Connect the power supply to the scale. 

Battery discharged. 

Connect  the  power  supply  to  the  mains, 
charge the battery. 

No  battery  (not  installed  or 
installed incorrectly). 

Check  if  the  battery  is  installed  correctly 
(polarization). 

The  scale  switches 
off automatically. 

<7.4.t1>  parameter  set  to 
value  enforcing  scale  shut-
down  after  particular  time 
interval. 

Go  to  <P7.Othr>  menu,  set  <7.4.t1> 
parameter to 'nonE' value. 

During  the  start-up, 
message 

'LH' 

is 

displayed. 

Weighing  pan  loaded  during 
the start-up. 

Unload the weighing pan. Zero indication 
is displayed. 

Communication  with 
the 

computer 

not 

established. 

Incorrect  computer  port  set  in 
parameter <5.1.1.Prt>. 

Enter < P5.ducE / 5.1.PC> submenu and 
set correct <5.1.1.Prt> parameter value. 

Incorrect 

transmission 

parameters  for  the  selected 
computer port. 

Enter  <P4.Conn>  menu  and  set  correct 
transmission  parameters for  the selected 
computer port. 

Incorrect

 

printout  frequency 

for continuous transmission. 

Enter < P5.ducE / 5.1.PC> submenu and 
set correct <5.1.3.Int> parameter value. 

No  printout 

on 

scale-connected 
printer. 

Incorrect  printer  port  set  in 
<5.2.1.Prt> parameter. 

Enter < P5.ducE / 5.2.Prtr> submenu and 
set correct <5.2.1.Prt> parameter value. 

Incorrect 

transmission 

parameters  for  the  selected 
printer port. 

Enter  <P4.Conn>  menu  and  set  correct 
transmission  parameters for  the selected 
printer port. 

No 

variable 

declared 

in 

weighing printout project. 

Enter <P6.Prnt  / 6.2.GLP> submenu  and 
declare variables that are to be printed. 

Communication  with 
the  additional  display 
not established.. 

Incorrect  additional  display 
port 

set 

in 

<5.3.1.Prt> 

parameter. 

Enter  <  P5.ducE  /  5.3.AdSP>  submenu 
and  set  correct  <5.3.1.Prt>  parameter 
value. 

Incorrect 

transmission 

parameters  for  the  selected 
computer port. 

Enter  <P4.Conn>  menu  and  set  correct 
transmission  parameters for  the selected 
additional display port. 

Displayed  mass  unit 
does  not  comply  with 
the scale data plate. 

Changed  scale  start  unit  in 
<9.1.UnSt> parameter. 

Enter <P9.Unit / 9.1.UnSt> submenu and 
set  unit  complying  with  the  scale  data 
plate. 

Changed 

custom 

unit 

in 

<9.2.Unin> parameter. 

Enter <P9.Unit  / 9.2.Unin> submenu  and 
set  unit  complying  with  the  scale  data 
plate. 

Содержание WIC/A1/C/2

Страница 1: ......

Страница 2: ...2 OCTOBER 2020...

Страница 3: ...ely unplug the device from the mains Scales to be decommissioned must be decommissioned in accordance with valid legal regulations Do not let battery discharge in case of prolonged storage of the devi...

Страница 4: ...rrangement 10 6 3 RS232 Connectors 10 6 4 Inputs Outputs 10 6 4 1 Technical Specifications 11 6 4 2 I O Schematic Diagrams 11 6 5 Operation panel 12 7 INSTALLATION 13 7 1 Unpacking and Installation 13...

Страница 5: ...can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a resi...

Страница 6: ...ificate H Contact with the central authorized service 48 48 386 63 30 3 MAINTENANCE ACTIVITIES In order to ensure safety in the course of cleaning it is necessary to disconnect the device from the mai...

Страница 7: ...cleansers from the panes prior their reinstallation After preliminary cleaning process stage rinse the pane using running water first distilled next Avoid drying the panes either using paper towel or...

Страница 8: ...ed component must be replaced or repaired by RADWAG service directly In case of any problems with correct operation of the scale contact the closest manufacturer s service point In case of defects del...

Страница 9: ...9 6 MECHANICAL DESIGN 6 1 Dimensions WLC A1 WLC A1 C 2 precision scale WLC A2 WLC A2 C 2 precision scale...

Страница 10: ...vice 6 3 RS232 Connectors RS232 1 connector DB9 M male Pin2 RxD Pin3 TxD Pin4 5 VDC Pin5 GND RS232 2 connector DSUB15 F female Pin8 TxD2 Pin9 5 VDC Pin10 GNDRS Pin13 RxD2 6 4 Inputs Outputs Standard C...

Страница 11: ...cal Specifications Output parameters Output quantity 4 Output type Solid state relay Cable cross section 0 14 0 5 mm 2 Maximum output current 0 5 A DC Maximum output voltage 30 VDC Input parameters In...

Страница 12: ...panel of WLC A1 C 2 WLC A2 C 2 precision scales Keys Press to switch the weighing device on off hold the key for about 1 second Function key press to change the working mode Press to send the weighin...

Страница 13: ...ON 7 1 Unpacking and Installation Unpack the device and place it on a flat and even surface Keep it far away from any sources of heat Install the weighing pan follow the below figures WLC A1 WLC A1 C...

Страница 14: ...it following section 7 1 of this user manual 2 Put the scale right side down 3 Remove the hole plug 4 Mount the hook and put the scale bottom side down 7 3 Levelling To level the weighing instrument t...

Страница 15: ...eading unit next displayed reading unit is conditioned by scale type In case the weighing result is not zero after indication stabilisation press key 7 5 Warm Up Time For correct scale operation the w...

Страница 16: ...ive status is displayed on the screen for 2s 80 Battery power supply Battery power given in CHArGE Battery charge in progress The device is connected to the power supply charging the battery Err5 Batt...

Страница 17: ...ess key repeatedly 9 WEIGHING Load the weighing pan Read the result when stability marker is displayed To assure long term operation and correct mass measurements follow the rules presented below Load...

Страница 18: ...18 10 DIAGRAMS OF CONNECTION CABLES scale computer cable scale printer cable EPSON I O cable...

Страница 19: ...computer port Incorrect printout frequency for continuous transmission Enter P5 ducE 5 1 PC submenu and set correct 5 1 3 Int parameter value No printout on a scale connected printer Incorrect printer...

Страница 20: ...art mass error indication out of range 10 of start mass H i Display range of total mass on scale display exceeded in Totalizing mode u L o Too low battery charge The scale is about to shut down E r r...

Страница 21: ...21...