- 35 -
1. Enter main menu, to do it press
Setup
key.
2. Enter < Profiles> submenu.
3. Press and hold profile that is to be copied.
4. Menu referring to a profile record is displayed.
5. Select <Copy> parameter.
6. A new <Copy
name
> profile is created, all new profile setting and base profile settings are
likewise.
7. Modify necessary data: (name, etc.).
Adding a new profile
Procedure:
1. Enter main menu, to do it press
Setup
key.
2. Enter < Profiles> submenu.
3. Press
key, message box: <Create new record?> is displayed.
4. Press
key to confirm. The software automatically adds a new record and enters its
editing mode.
Deleting a profile
Procedure:
1. Enter main menu, to do it press
Setup
key.
2. Enter < Profiles> submenu.
3. Press and hold profile that is to be deleted.
4. Menu referring to a profile record is displayed.
5. Select <Delete> parameter.
6. Message box: <Confirm to delete> is displayed.
7. Press
key to confirm, the profile gets deleted.
CAUTION!
In order to operate profiles it is necessary to log in as an Administrator.
13.2. PROFILE STRUCTURE
Each profile contains the following entries:
Settings
Enter this menu to set an individual profile name (a order of alphanumeric
characters), and to declaring a default working mode (the selected mode is
activated as default on profile selecting).
Working modes
Contains the following submenu:
•
Additional setting of a working mode
•
Function keys
•
Displayed information
•
Printouts
Readout
Contains the following submenu:
•
Filter
•
Value release
•
Autozero
•
Autozero: Dosing
•
Last digit
Units
Enter this menu to declare the start unit, the supplementary unit, 2 custom
units and to enter the value of gravitational acceleration force in balance’s
operation place.