- 108 -
Error value [- T]
Max permissible low limit error T (-T), given in unit set for
a product. Measurements lower than value of Qn-T are
regarded as invalid.
Error value [+ T]
Max permissible high limit error T (+T), given in unit set
for a product. Measurements higher than value of Qn+T
are regarded as invalid.
Disqualifying quantity of
samples [Qn-2T]
Number of minus errors (–2T) of a tested sample which
disqualifies control.
Disqualifying quantity of
samples [Qn+2T]
Number of minus errors (+2T) of a tested sample which
disqualifies control.
Disqualifying quantity of
samples [Qn-T]
Number of minus errors (–T) of a tested sample which
disqualifies control.
Disqualifying quantity of
samples [Qn+T]
Number of plus errors (+T) of a tested sample which
disqualifies control.
Average limit
Mode for determination of average limit value (constant or
automatic).
Average limit -
Average limit value (minus) for a tested sample (value of
average limit as “constant”).
Average limit +
Average limit value (plus) for a tested sample (value of
average limit as “constant”).
Coefficient value [-Wk]
Multiplier of standard deviation for value of average limit
(minus) calculated automatically.
Coefficient value [+Wk]
Multiplier of standard deviation for value of average limit
(plus) calculated automatically.
28.5. CONTROL START PROCEDURE
Procedure:
1. Log in, make sure your access level allows you to carry out control (for more information
read section on logging of this user manual).
2. Select product with correctly entered control data.
3. Enter weighing mode parameters:
Содержание UYA 4Y