- 105 -
28. PREPACKAGED GOODS CONTROLS
(disabled for standard balance versions)
<
Prepacked Goods Controls
> is a working mode enabling realizing prepackaged goods
controls (single stand control or multi-stand control) supported by a database containing products
and users. A control initiated from a balance is automatically finished after a pre-set number of
packages (samples) has been controlled.
Balances may connect to
E2R SYSTEM
software and form a multi-stand system (network). Each
balance is an independent weighing stand and information on control progress is transmitted
continuously to software. The software enables collection of real-time data from each connected
balance. Control may be activated either by means of a balance or the software.
Collected data allow you to assess quality of produced prepacked goods and their conformity with:
•
regulations concerning prepacked goods control of a given country;
•
a company quality control system (internal control).
Control cycle scheme:
•
Select a user.
•
Select a product.
•
Start control.
•
Acquire weighing records.
•
Automatically switch off the control after a determined number of packages (sample) has been
weighed.
•
Print a report on completed control.
CAUTION!
For detailed instruction on how to establish connection with <
E2R System> go to section 32.1 ‘Computer’
of this user manual.
28.1. PREPACKED GOODS CONTROL MODE START-UP
Procedure:
1. Go to balance home screen, press
button located in the top bar, <Working modes>
submenu opens.
2. Select
<
Prepacked Goods Controls>
mode, Prepacked Goods Controls home screen is
displayed.
3.
<Start control>
message is visible in
the bottom bar along with
button, press it to start the
control.