14. SCALE PARAMETERS
Users can set the scale according to the ambient conditions (filtering level)
or own needs (autozero) and set the LO threshold for minimum load that
enables operation of some functions. This parameters are placed in
<
Weighing>
.
In order to enter submenu
<
Weighing>,
press
and then
:
„
Weighing”
.
Caution:
Weighing parameters are directly related to a specific weighing platform,
so at the beginning the weighing platform should be selected for which
we want to set parameters.
Inventory of scale parameters:
Median Filter
Filter
Autozero
LO Threshold
Last digit
14.1. Median filter
The median filter is intended for eliminating short-lasting mechanical shocks.
Procedure:
•
Enter
<
Weighing>
according to ch. 14 of the manual, select
<
Median Filter>
and then set an appropriate value.
Accessible settings:
None
- median filter is off
0.5, 1, 1.5, 2, 2.5
- filtering level to choose