Enter requested correction and/or press
key to confirm.
<Correction>
edit box is displayed automatically.
Enter requested correction and/or press
key to confirm.
<Corrections>
submenu is displayed automatically.
9.7.3. Deleting Linearity
Enter
<Factory/Section linearity>
submenu.
Select
<Delete>
parameter. Message
<Delete?>
is displayed.
Press
key to confirm.
9.8. Gravitational Coefficient
The function of gravity correction compensates changes of gravity force being
a result of different latitude. It allows to carry out correct scale
calibration/adjustment away from the point of subsequent use. The gravity
correction value must be entered with reference to tables prepared by „Radwag
Balances and Scales” or calculated using the below formula:
kal.
uzy t.
g
g
Gcor
Correction value ranges between 0.90000 ÷ 1.99999.
If the weighing instrument is calibrated in the place of use,
then the value of <GCOR coefficient> parameter must be
1.00000. If the weighing instrument is calibrated away from
the place of use (longitudinal change) the value must be
corrected.