![RADWAG PUE HX7 Скачать руководство пользователя страница 29](http://html2.mh-extra.com/html/radwag/pue-hx7/pue-hx7_user-manual_3386998029.webp)
29
9.6. Factory Adjustment
9.6.1. External Adjustment
Enter
<Factory/Adjustment>
submenu.
Select
<External adjustment>
parameter. Message
<Remove weight>
is displayed.
Unload the weighing pan and press
key to confirm.
Message
<Adjustment. Please wait…>
is displayed.
Upon completed start mass determination procedure, message
<Put weight xxx>
is displayed (where: xxx – adjustment weight mass).
Load the weighing pan with a required adjustment weight and press
key to confirm. Message
<Adjustment. Please wait…>
is displayed.
Upon adjustment completion, message
<Remove weight>
is displayed.
Unload the weighing pan, the weighing device displays the
<Adjustment>
submenu.
The adjustment of additional platform is performed likewise.
9.6.2. Start Mass Determination
Enter
<Factory/Adjustment>
submenu.
Select
<Start mass determination>
parameter. Message
<Remove
weight>
is displayed.
Unload the weighing pan and press
key to confirm.
Message
<Start mass de
termination. Please wait…>
is displayed.
Upon completed determination process, the weighing device displays the
<Adjustment>
submenu.
Process of determination of start mass for an additional
platform calibration is analogous.