15
8.2. Factory Parameters List
No.
Name
Value
Description
P0.
FAct
-
Factory parameters
0.1.
Glob
-
Global parameters
0.1.1. duu
-
Defining the weighing device.
0.1.2. FAc
Serial number.
0.1.3. tYP
1, 2, 4, 6, 7, 8, 9, 12
Weighing device type:
1
- WLC/A2;
2
- WLC/F, WLC/C2,
4
– WTC,
6
– Medical scale,
7
- Medical scale
(BMI disabled),
8
– PUE C315;
9
– PUE H315;
12
– WLC/C/2.
0.1.4. Gcor
0.9 - 1.1
Gravity correction factor.
0.1.7. tSc
SLA, nInnH, no
Battery selection.
0.1.8. CSt
nonE, d, A, V, b, SP, SC,
nt
Declaring the customer:
nonE
- none,
d
- KERN,
A
- ADEMI,
V
- VWR,
b
- BOECO,
SP
- Spectrum,
SC
– Schuller,
nt
– NEW TECH.
0.1.9. rtc
-
RTC synchronisation.
0.1.A. ntE
YES, no
US market metrological requirements
activation.
0.2.
nnG
-
Metrology
0.2.1. A/d
-
Converter's divisions.
0.2.2. Uni
g, kg, lb
Adjustment unit.
0.2.3. du1
0.0001, 0.0002, 0.0005,
0.001, 0.002, 0.005, 0.01,
0.02, 0.05, 0.1, 0.2, 0.5, 1,
2, 5, 10, 20, 50
Range 1 reading unit.
0.2.4. dE1
no, 0.001, 0.01, 0.1, 1, 2,
5
Range 1 verification unit;
no
- non-
verified balance/scale.
0.2.5. du2
0.0001, 0.0002, 0.0005,
0.001, 0.002, 0.005, 0.01,
0.02, 0.05, 0.1, 0.2, 0.5,
1, 2, 5, 10, 20, 50
Range 2 reading unit.
0.2.6. dE2
no, 0.001, 0.01, 0.1, 1, 2,
5
Range 2 verification unit;
no
- non-
verified balance/scale.
0.2.7. Ful
-
Maximum range + overload.
0.2.8. rn2
-
Range switching point.
0.2.9. uuE
-
External adjustment weight mass.
0.2.A
uui
-
Internal adjustment weight mass.
„0”
- internal adjustment disabled.