Procedure: 

 

  Enter 

<Factory/Section  linearity>

  submenu  and  select 

<Determine>

 

parameter. Message 

<Continue?>

 is displayed. 

  Press 

 key to confirm. 

<Mass>

 edit box is displayed. 

  Unload the weighing pan. 

  Enter  the  required  mass  value  (the  first  linearity  correction  point)  and 

press 

  key  to  confirm.  Message 

<Put  weight  xxx>

  is  displayed 

(where: xxx 

– entered mass value) is displayed. 

  Load the weighing pan remembering that the load mass must equal weight 

value  specified  for  the  first  linearity  correction  point  (it  is  possible  to 
preview the weighing result in the displayed window).  

  Upon  result  stabilization,  press 

  key  to  confirm. 

<Mass>

  edit  box  

is displayed again (second linearity correction point). 

  Linearity correction process for the 2nd point is analogous to the linearity 

correction process for the 1st point. 

  Upon  declaring  requested  quantity  of  linearity  correction  points,  press 

key when 

<Mass>

 edit box is displayed again. 

 

<Section linearity>

 submenu is displayed automatically. 

 

 

Upon pressing 

 

key to confirm zero value for linearity 

correction,

 

message: <Value too low> is displayed. 

 
 

9.6.2. Corrections 

 

Editing corrections for linearity correction points. 

 

Procedure: 

 

  Enter 

<Factory/Section linearity>

 submenu.  

  Select 

<Corrections>

  parameter. A  list  of  determined  linearity correction 

points  is  displayed.  The  list  contains  previously  determined  mass  values 
for each correction point and deviations determined in reading units. 

  Select requested position, 

<Mass>

 edit box is displayed. 

  Enter  requested  correction  and/or  press 

  key  to  confirm. 

<Correction>

 edit box is displayed automatically. 

  Enter  requested  correction  and/or  press 

  key  to  confirm. 

<Corrections>

 submenu is displayed automatically. 

Содержание PUE C32

Страница 1: ......

Страница 2: ...2 AUGUST 2020...

Страница 3: ...damage immediately unplug the device from the mains Scales to be decommissioned must be decommissioned in accordance with valid legal regulations Do not let battery discharge in case of prolonged stor...

Страница 4: ...4 Pictograms 13 7 OPERATING THE MENU 13 7 1 Entering the Menu 13 7 2 Menu Keys 14 7 3 Entering Numbers Text 15 7 3 1 Numerical Box 15 7 3 2 Text Box 16 7 3 3 Diacritical Sign Table 18 7 3 4 Special Si...

Страница 5: ...d with the assistance of the manufacturer and the user representatives C RADWAG does not bear any responsibility for damage or losses resulting from unauthorized or inadequate performing of production...

Страница 6: ...pecial cleaning agents based on a mixture of aliphatic hydrocarbons that do not dissolve plastics Before using the cleanser for all surfaces we recommend carrying out tests Do not use cleansers contai...

Страница 7: ...B connector 4 3 RS232 Connectors RS232 DB9 M connector male front Pin2 RxD Pin3 TxD Pin4 5VDC Pin5 GND RS232 DSUB15 F connector female front Pin8 TxD2 Pin9 5VDC Pin10 GND Pin13 RxD2 4 4 Inputs Outputs...

Страница 8: ...Technical Specifications Output parameters Output quantity 4 Output type Solid state relay Cable cross section 0 14 0 5mm 2 Maximum output current 0 5A DC Maximum output voltage 30VDC Input parameters...

Страница 9: ...ess to log in Press to cancel the message Press to zero the scale Press to tare the scale Press to change the weighing unit Press to confirm the message Press to confirm the weighing result PRINT Pres...

Страница 10: ...nnections 4 or 6 wires shield Load cell excitation voltage 5V DC Maximum signal gain 19 5 mV Display 5 colour graphic display IP rating IP43 Housing ABS plastic Maximum number of connected platforms 1...

Страница 11: ...battery status or informs that battery charging is in progress pictogram displayed in a cyclic manner battery charging pictogram displayed continuously battery charged within 75 100 range of permissi...

Страница 12: ...re the home screen is to be found in PUE C32 Indicator Software Manual 6 1 Top Bar The top bar displays the following information Working mode name and symbol Device name Wireless communication on Com...

Страница 13: ...n customize the screen 6 4 Pictograms The pictograms assigned to operation panel keys are to be found underneath the workspace You can define on screen pictograms individually for each working mode 7...

Страница 14: ...n editing numeric and text values Press to select change working mode Press to confirm accept modifications Press to go back Press to discard entering parameter modifications Press to go up the menu P...

Страница 15: ...ariable value etc Button functions change depending on the edit box type 7 3 1 Numerical Box Where Digit mode To change keyboard working mode press key Keyboard working mode options upper case charact...

Страница 16: ...nter digit 6 Press to enter digit 7 Press to enter digit 8 Press to enter digit 9 Press to enter digit 0 Press to enter sign dot long press Press to delete one character Press to exit the edit box con...

Страница 17: ...t Keys Press to enter Press to enter a b c Press to move the cursor to the left long press Press to enter d e f Press to move the cursor up long press Press to enter g h i Press to move the cursor to...

Страница 18: ...s characteristic for a particular interface language are automatically added to the table when the given language gets selected Diacritical sign table Polish Diacritical sign table English German Fren...

Страница 19: ...press key repeatedly keep pressing the key until you see the home screen press key the home screen is displayed immediately 8 INSTALLER INSTRUCTION The PUE C32 indicator serves as basis of load cell...

Страница 20: ...IELD See section 9 3 REF SENSE JP3 not soldered REF SENSE JP4 not soldered IN OUTPUT IN OUTPUT 5V INPUT AGND INPUT 8 2 Connecting 4 Wire Load Cell Connect 4 wire load cell to the main board following...

Страница 21: ...ndicator on a post etc DO NOT connect E E Soldering spot on the main board and on the additional A D converter boards 9 FACTORY SETUP To access and modify factory settings and parameters that are made...

Страница 22: ...cale type PUE C32 PUE C32 C32 WLX Setting scale type For PUE C32 value only weighing indicator name is displayed metrological data is not Scale name Special name and or name of RADWAG distributor s de...

Страница 23: ...02 0 005 0 01 0 02 0 05 0 1 0 2 0 5 1 2 5 10 20 50 Verification units for Range 1 none value unverified version Reading unit Range 2 0 001 0 0000001 50 Reading units for Range 2 Verification unit Rang...

Страница 24: ...hreshold parameter Start mass threshold 10 10 90 Start mass value range in Adjustment Weighing instrument adjustment menu External adjustment Scale adjustment refer to section 9 5 1 Start Mass Determi...

Страница 25: ...ed where xxx adjustment weight mass Load the weighing pan with a required adjustment weight and press key to confirm Message Adjustment Please wait is displayed Upon adjustment completion message Remo...

Страница 26: ...analogous to the linearity correction process for the 1st point Upon declaring requested quantity of linearity correction points press key when Mass edit box is displayed again Section linearity subme...

Страница 27: ...from the point of subsequent use The gravity correction value must be entered with reference to tables prepared by Radwag Balances and Scales or calculated using the below formula kal uzy t g g Gcor C...

Страница 28: ...28 Scale printer cable ZEBRA Scale printer cable EPSON I O cable Scale Ethernet cable is a standard network cable terminated with RJ45 connectors on both ends...

Страница 29: ...29...

Отзывы: