Programmable key assigned to an on-screen button.
key
– long press to get scale info.
Programmable key assigned to an on-screen button.
Programmable key assigned to an on-screen button.
Programmable key assigned to an on-screen button.
Programmable key assigned to an on-screen button.
4.6. Technical Specifications
Maximum quantity of verification units
6000
Maximum impedance of load cell
1200 Ω
Minimum impedance of load cell
50 Ω
Maximum voltage per verification unit
3.25 µV
Minimum voltage per verification unit
0.4 µV
Load cell connections
4 or 6 wires + shield
Load cell excitation voltage
5V DC
Maximum signal gain
19.5 mV
Display
5” colour graphic display
IP rating
IP43
Housing
ABS plastic
Maximum number of connected platforms
1
Power supply
10 ÷ 16VDC
Operating temperature
0°C ÷ 40°C
Communication interfaces
2xRS232, USB A, USB B, Ethernet,
wireless communication, 4I/O (option)
Multi range option
YES
5. INDICATOR INSTALLATION
5.1. Unpacking and Installation
A. Take the indicator out of the packaging.
B. Connect the indicator to a weighing platform, next place the weighing
device on a flat and even surface. Keep it far away from any sources of
heat.
C. To level the weighing instrument turn its feet. Keep turning the feet until
the air bubble takes central position: