26. STATISTICAL QUALITY CONTROL- SQC
<
Statistical Quality Control
> is a working mode designed to support control of product
packing process. It enables detecting surplus or shortfall of product in a packaging.
If the weighing results are stored in the database, then trend analysis can be carried out and
displayed in a form of a graph.
The software provides the possibility of controlling a batch comprising up to 1000 samples.
Each completed control process is saved to the SQC database, and its results are ready for
previewing at any time. Each series has a determined set of parameters that are stored in
the database: maximum, minimum, standard deviation and average value for each batch,
etc.
SQC control process can be carried out manually (by pressing <PRINT> button for each
measurement) or automatically (automatically carried out record of stable measurements).
The balances feature module for statistical control processes. It is supported by a database
containing list of products with error values (<T2->; <T1->; <T1+> and <T2+>) declared for
each product.
The control process is initiated automatically and completed on controlling a required
quantity of samples. As a user you define how many samples are to be controlled, to do it go
to SQC settings.
After completed control process a final report can be generated and/or printed. Data on a
control is also automatically saved to the Database of SQC Reports.
Control process:
•
operator selection,
•
product selection,
•
control start,
•
acquiring weighing records,
•
control finish (carried out automatically when predefined number of samples is weighed -
batch),
•
control report print.
26.1. SQC Mode Start Up
In order to activate SQC mode you need to be assigned with permission level required for
carrying out control process.
CAUTION:
1. Starting a control requires <Operator> permissions level at least. If a logged user or
an anonymous user’ permissions level is <Guest>
then on control start, message
<Unauthorized access> is displayed.
2. For information on logging procedure read section 8 of this user manual, for
information on how to grant permissions read section 12 of this user manual.
•
Set
<
SQC>
mode parameters. For detailed information read section 26.2.
•
Select a product that is to be controlled (press <
> button), mind to select a product
with correctly entered control data,
List of data to be entered prior control start
Содержание MAS 1. Y
Страница 1: ...WEIGHING MODULES MAS Y MPS Y MAS 1 Y USER MANUAL IMMU 25 01 08 18 EN www radwag com...
Страница 9: ...9 1 GENERAL INFORMATION 1 1 Dimensions MAS Module MPS Module...
Страница 10: ...10 MAS 1 Module...
Страница 200: ...200...
Страница 201: ...201...