- 88 -
8. Keep following the prompts until the calibration process is completed.
INFO field provides you with information on the on-going process (average volume,
error values).
9. Upon confirmation of the last measurement a window with ambient conditions
parameters is displayed again. Enter the parameters and press <
Accept>
button to confirm.
10.
A report is generated automatically, the report is printed and recorded in pipettes
calibration reports database (ambient conditions and water temperature values given
on a report are average values from the process);
11.
Press <
> button located in the bottom bar. For multichannel pipette you are
asked whether the process shall be continued for remaining channels. Upon
confirmation the balance carries out calibration process for the following channels.
The settings remain unchanged.
12.
You can start the following procedure for the same pipette or select another pipette
form pipettes database.
24.6. Performed Pipettes Calibrations Report
Upon completed calibration process a report is generated. The report is saved to the <
Pipette calibration reports
> database. The files are named by date and time (hour) of
calibration process performance.
Report example:
----------- Pipettes Calibration -----------
Operator Kowalski
Customer Nowak
Pipette p901\1k
Serial no. 7777
Channels quantity 1
Channel no. 1
Measurements quantity 10
Operation with adherence to ISO 8655 Yes
Start date 2012.03.15 07:50:44
End date 2012.03.15 07:54:34
Water temperature 22.15 °C
Temperature 21 °C
Humidity 48 %
Pressure 1005 hPa
Z coefficient 1.00328
------- Tested volume: 1000 µl -------
1 0.998 g 1000.82389 µl
2 0.998 g 1000.82389 µl
3 0.998 g 1000.82389 µl
4 0.998 g 1000.82389 µl
5 0.998 g 1000.82389 µl
.
.
Average volume 1000.82389 µl
Average [%] 100.08 %
Systematic error 0.82389 µl
Systematic error[%] 0.08239 %
Permissible error ± 16 µl
Random error 0 µl
Random error [%] 0 %
Permissible error ± 6 µl
Status Positive
----------------------------------------
Signature
........................................